Py/TD Thermal Pyrolysis Section
|
|
Thermal Pyrolysis Temperature Control Range
|
Above room temperature, 5°C - 600°C
|
|
Temperature Control Precision
|
±0.1℃
|
|
Heating Rate
|
<200℃/min
|
|
Thermal Pyrolysis Temperature Control Method
|
5-stage programmed temperature rise control
|
|
Scrubbing Flow Rate
|
0-500mL/min
|
|
Interface Temperature Range
|
Above room temperature, 5°C - 600°C
|
|
Injection Port Parameters
|
|
Maximum Installation Quantity
|
Up to 3 injection ports can be installed
|
|
Optional Injection Ports
|
PIPP (Packed Insert Port with PTFE Cap), Standard and Inert Flow Path Split/Splitless Capillary Injection Port (S/SL), Cold Tip Injection Port (COC)
|
|
Injection Port Temperature
|
Above room temperature, 5°C - 450°C
|
|
Tip Pressure Setting Range
|
0 - 100 psi, Control Setting Precision: 0.01 psi
|
|
Total Flow Rate Setting Range
|
0 - 300 mL/min, Setting Precision: 0.01 mL/min
|
|
Gas Control Method
|
Constant Pressure, Constant Flow Rate
|
|
Column Temperature Box Parameters
|
|
Furnace dimensions
|
28×30×18 cm. It can accommodate two 105 m × 0.530 mm inner diameter capillary columns or two 10-foot glass packed columns (Coil diameter 9 inches, 1/4 inch outer diameter) or two 20-foot long stainless steel packed columns (1/8 inch outer diameter).
|
|
Temperature range
|
Room temperature 5℃ to 450℃
|
|
Programmed temperature rise
|
16 stages of programmable temperature rise can be set
|
|
Temperature control accuracy
|
0.1℃
|
|
Maximum heating rate
|
120℃/min
|
|
Column oven cooling speed
|
300℃ to 50℃ is less than 360 seconds
|
|
Longest single method running time
|
999.99min
|
|
Heating areas
|
6 independent controlled heating zones, auxiliary heating zone maximum operating temperature: 300℃
|
|
Protection function
|
Automatic protection function for column oven temperature
|
|
Detector Parameters
|
|
Configuration method
|
Multiple independent temperature-controlled detectors can be equipped. All detectors include an electronic flow control module.
|
|
Optional detectors
|
FID、TCD、u-ECD 、FPD、PDHID、MS
|
|
Maximum operating temperature of FID
|
450°C
|
|
Detection limit of FID
|
≤ 2.0 pg C/s (n-hexadecane)
|
|
Baseline drift (30 minutes)
|
≤ 3×10⁻¹³ A
|
|
Baseline noise
|
≤ 5×10⁻¹⁴ A
|
|
Dynamic linear range
|
≥10⁷
|
|
Data acquisition frequency
|
Up to 500Hz
|
|
Software Parameters
|
|
Compatible Systems
|
Supports Windows 7/10/11, 32-bit graphical user interface
|
|
Core Control Functions
|
Can quickly set parameters such as injection port, detector, column temperature box heating program, chromatographic column conditions, and valve events
|
|
Data Processing Functions
|
Double-click to automatically identify chromatographic peak retention time, multi-spectrum comparison, repeatability analysis, automatic sequence acquisition, automatic integration correction, and report output
|
|
Qualitative and Quantitative Functions
|
Qualitative: Supports multiple relative retention times and grouping; Quantitative: Area percentage method, corrected area percentage method, internal standard method, external standard method, can manually create working curves
|
|
System Auxiliary Functions
|
GC automatic stop/automatic start, system self-diagnosis, status recording, Chinese and English selectable
|
|
Data Communication
|
Ethernet (LAN), remotely start/stop operation start/end, TCP/CP protocol, stable and distortion-free transmission
|
|
Overall Specifications
|
|
Overall dimensions
|
570mm×650mm×550mm
|
|
Power supply
|
220V±10%(50Hz±0.5Hz)
|
|
Power
|
3000W
|